Aufprallprüfung BrustimplantateFor the impact test according to ISO 14607:2024 Annex C.2, INNOPROOF has developed and built a custom drop rig that ensures a standard-compliant test. Depending on the mass and projection height of the breast implant, the required drop height is calculated. A weight of 4.4 kg is held by an electromagnet, and the drop height is set using a measuring device. The test specimen is placed centrally on the test table. When the electromagnet is switched off, the weight is released, initiating the impact.

After the impact test, the breast implant undergoes a visual inspection. To pass the test, no damage to the shell may be visible.