<?xml version="1.0" encoding="UTF-8"?><rss version="2.0"
	xmlns:content="http://purl.org/rss/1.0/modules/content/"
	xmlns:wfw="http://wellformedweb.org/CommentAPI/"
	xmlns:dc="http://purl.org/dc/elements/1.1/"
	xmlns:atom="http://www.w3.org/2005/Atom"
	xmlns:sy="http://purl.org/rss/1.0/modules/syndication/"
	xmlns:slash="http://purl.org/rss/1.0/modules/slash/"
	>

<channel>
	<title>INNOPROOF GmbH</title>
	<atom:link href="https://innoproof.de/en/feed/" rel="self" type="application/rss+xml" />
	<link>https://innoproof.de/en/</link>
	<description></description>
	<lastBuildDate>Tue, 29 Sep 2026 10:37:31 +0000</lastBuildDate>
	<language>en-GB</language>
	<sy:updatePeriod>
	hourly	</sy:updatePeriod>
	<sy:updateFrequency>
	1	</sy:updateFrequency>
	<generator>https://wordpress.org/?v=7.1.2</generator>

<image>
	<url>https://innoproof.de/wp-content/uploads/2025/11/cropped-Innoproof_favicon-32x32.webp</url>
	<title>INNOPROOF GmbH</title>
	<link>https://innoproof.de/en/</link>
	<width>32</width>
	<height>32</height>
</image> 
	<item>
		<title>DIN EN 868-5:2019 Peel Test of Flexible Packaging</title>
		<link>https://innoproof.de/en/din-en-868-52019-peel-test-of-flexible-packaging/</link>
		
		<dc:creator><![CDATA[Admin21]]></dc:creator>
		<pubDate>Tue, 29 Sep 2026 10:33:32 +0000</pubDate>
				<category><![CDATA[Testing Services]]></category>
		<category><![CDATA[Mechanical testing of packaging for medical devices]]></category>
		<guid isPermaLink="false">https://innoproof.de/?p=1758</guid>

					<description><![CDATA[<p>The peel test according to DIN EN 868-5 (Annex D) is a standardized test method for determining the opening behavior (peelability) of sealable pouches and reels for the sterilization of medical devices.<br />
The goal is to ensure that the seal strength is sufficient to keep the product hermetically sealed during transport and sterilization, while also allowing it to be peeled open cleanly without residue (without fiber shedding or tearing of the film).</p>
<p>The separation process is evaluated visually:</p>
<p> 	Fiber-free opening: No loose fibers must form when separating paper or nonwoven layers, as these could fall onto the sterile product and cause contamination.<br />
 	No delamination/tearing: The film must not tear or split into multiple layers.<br />
 	Uniformity: The seal must open continuously without jerky force spikes (stick-slip effect).</p>
<p>Der Beitrag <a href="https://innoproof.de/en/din-en-868-52019-peel-test-of-flexible-packaging/">DIN EN 868-5:2019 Peel Test of Flexible Packaging</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></description>
										<content:encoded><![CDATA[<p>The peel test according to DIN EN 868-5 (Annex D) is a standardized test method for determining the opening behavior (peelability) of sealable pouches and reels for the sterilization of medical devices.</p>
<p>The goal is to ensure that the seal strength is sufficient to keep the product hermetically sealed during transport and sterilization, while also allowing it to be peeled open cleanly without residue (without fiber shedding or tearing of the film).</p>
<p>The separation process is evaluated visually:</p>
<ul>
<li>Fiber-free opening: No loose fibers must form when separating paper or nonwoven layers, as these could fall onto the sterile product and cause contamination.</li>
<li>No delamination/tearing: The film must not tear or split into multiple layers.</li>
<li>Uniformity: The seal must open continuously without jerky force spikes (stick-slip effect).</li>
</ul>
<p>Der Beitrag <a href="https://innoproof.de/en/din-en-868-52019-peel-test-of-flexible-packaging/">DIN EN 868-5:2019 Peel Test of Flexible Packaging</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></content:encoded>
					
		
		
			</item>
		<item>
		<title>ASTM F1980: Accelerated aging of sterile barrier systems and medical devices</title>
		<link>https://innoproof.de/en/astm-f1980-accelerated-aging-of-sterile-barrier-systems-and-medical-devices/</link>
		
		<dc:creator><![CDATA[Admin21]]></dc:creator>
		<pubDate>Tue, 29 Sep 2026 10:28:27 +0000</pubDate>
				<category><![CDATA[Testing Services]]></category>
		<category><![CDATA[Mechanical testing of packaging for medical devices]]></category>
		<guid isPermaLink="false">https://innoproof.de/?p=1753</guid>

					<description><![CDATA[<p>The ASTM F1980 standard (“Standard Guide for Accelerated Aging of Sterile Barrier Systems and Medical Devices”) provides guidelines for conducting accelerated aging tests on sterile packaging and medical devices. Its objective is to rapidly demonstrate expiration dates (shelf life) and the long-term integrity of sterile barrier systems (e.g., pouches, blister packs) without having to wait for years of real-time aging.</p>
<p>The packaged products are stored at elevated temperatures (commonly between 50 °C and 60 °C) and controlled humidity levels to thermally accelerate physical degradation processes. The mathematical basis of accelerated aging is the Arrhenius equation. This is grounded in the model that a 10 °C temperature increase approximately doubles the rate of aging reactions. As a result, 12 months of real-time aging can often be simulated in just a few weeks within an environmental chamber.</p>
<p>Following the calculated test duration, the packaging is tested for parameters such as seal strength, integrity (sterile barrier protection), and material embrittlement, while the medical device itself is evaluated for functionality.</p>
<p>Der Beitrag <a href="https://innoproof.de/en/astm-f1980-accelerated-aging-of-sterile-barrier-systems-and-medical-devices/">ASTM F1980: Accelerated aging of sterile barrier systems and medical devices</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></description>
										<content:encoded><![CDATA[<p><img fetchpriority="high" decoding="async" class="alignright wp-image-1751 size-medium" src="https://innoproof.de/wp-content/uploads/ASTM-F1980-accelerated-aging-300x200.webp" alt="ASTM F1980 accelerated aging" width="300" height="200" srcset="https://innoproof.de/wp-content/uploads/ASTM-F1980-accelerated-aging-200x133.webp 200w, https://innoproof.de/wp-content/uploads/ASTM-F1980-accelerated-aging-300x200.webp 300w, https://innoproof.de/wp-content/uploads/ASTM-F1980-accelerated-aging-400x267.webp 400w, https://innoproof.de/wp-content/uploads/ASTM-F1980-accelerated-aging-768x512.webp 768w, https://innoproof.de/wp-content/uploads/ASTM-F1980-accelerated-aging.webp 900w" sizes="(max-width: 300px) 100vw, 300px" />The ASTM F1980 standard (“Standard Guide for Accelerated Aging of Sterile Barrier Systems and Medical Devices”) provides guidelines for conducting accelerated aging tests on sterile packaging and medical devices. Its objective is to rapidly demonstrate expiration dates (shelf life) and the long-term integrity of sterile barrier systems (e.g., pouches, blister packs) without having to wait for years of real-time aging.</p>
<p>The packaged products are stored at elevated temperatures (commonly between 50 °C and 60 °C) and controlled humidity levels to thermally accelerate physical degradation processes. The mathematical basis of accelerated aging is the Arrhenius equation. This is grounded in the model that a 10 °C temperature increase approximately doubles the rate of aging reactions. As a result, 12 months of real-time aging can often be simulated in just a few weeks within an environmental chamber.</p>
<p>Following the calculated test duration, the packaging is tested for parameters such as seal strength, integrity (sterile barrier protection), and material embrittlement, while the medical device itself is evaluated for functionality.</p>
<p>Der Beitrag <a href="https://innoproof.de/en/astm-f1980-accelerated-aging-of-sterile-barrier-systems-and-medical-devices/">ASTM F1980: Accelerated aging of sterile barrier systems and medical devices</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></content:encoded>
					
		
		
			</item>
		<item>
		<title>ASTM D6344: Concentrated Impacts to Transport Packages</title>
		<link>https://innoproof.de/en/astm-d6344-concentrated-impacts-to-transport-packages-2/</link>
		
		<dc:creator><![CDATA[Admin21]]></dc:creator>
		<pubDate>Tue, 29 Sep 2026 10:21:04 +0000</pubDate>
				<category><![CDATA[Testing Services]]></category>
		<category><![CDATA[Mechanical testing of packaging for medical devices]]></category>
		<guid isPermaLink="false">https://innoproof.de/?p=1748</guid>

					<description><![CDATA[<p>The ASTM D6344 standard (“Standard Test Method for Concentrated Impacts to Transport Packages”) describes a test method for determining the resistance of transport packaging to concentrated or point impacts. Its objective is to simulate localized impacts that occur during transport and handling—such as bumping against sharp loading edges, impact from falling cargo, or knocks from forklift forks.<br />
A test object in the form of a cylindrical steel rod with a hemispherical tip impacts specific weak points or surfaces of the packaging with a defined kinetic energy. In our custom-built test rig, this is carried out via a guided drop from a specified height.<br />
In contrast to the free-fall drop test (ASTM D5276), where the total weight of the packaging acts upon a surface, edge, or corner, the D6344 method tests the outer packaging’s resistance to penetration and denting against external, concentrated forces. The evaluation assesses whether the outer shell (e.g., corrugated cardboard, plastic, or wood) is punctured or severely indented and whether the contents were damaged as a result.</p>
<p>Der Beitrag <a href="https://innoproof.de/en/astm-d6344-concentrated-impacts-to-transport-packages-2/">ASTM D6344: Concentrated Impacts to Transport Packages</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></description>
										<content:encoded><![CDATA[<p>The ASTM D6344 standard (“Standard Test Method for Concentrated Impacts to Transport Packages”) describes a test method for determining the resistance of transport packaging to concentrated or point impacts. Its objective is to simulate localized impacts that occur during transport and handling—such as bumping against sharp loading edges, impact from falling cargo, or knocks from forklift forks.</p>
<p>A test object in the form of a cylindrical steel rod with a hemispherical tip impacts specific weak points or surfaces of the packaging with a defined kinetic energy. In our custom-built test rig, this is carried out via a guided drop from a specified height.</p>
<p>In contrast to the free-fall drop test (ASTM D5276), where the total weight of the packaging acts upon a surface, edge, or corner, the D6344 method tests the outer packaging’s resistance to penetration and denting against external, concentrated forces. The evaluation assesses whether the outer shell (e.g., corrugated cardboard, plastic, or wood) is punctured or severely indented and whether the contents were damaged as a result.</p>
<p>Der Beitrag <a href="https://innoproof.de/en/astm-d6344-concentrated-impacts-to-transport-packages-2/">ASTM D6344: Concentrated Impacts to Transport Packages</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></content:encoded>
					
		
		
			</item>
		<item>
		<title>ASTM D5276: Drop Test of Loaded Containers by Free Fall</title>
		<link>https://innoproof.de/en/astm-d5276-drop-test-of-loaded-containers-by-free-fall/</link>
		
		<dc:creator><![CDATA[Admin21]]></dc:creator>
		<pubDate>Tue, 29 Sep 2026 10:13:49 +0000</pubDate>
				<category><![CDATA[Testing Services]]></category>
		<category><![CDATA[Mechanical testing of packaging for medical devices]]></category>
		<guid isPermaLink="false">https://innoproof.de/?p=1740</guid>

					<description><![CDATA[<p>The ASTM D5276 standard (“Standard Test Method for Drop Test of Loaded Containers by Free Fall”) describes a standardized test procedure for determining the impact resistance of filled shipping containers. Its objective is to evaluate the ability of packaging (such as boxes, crates, drums, or sacks) to withstand drop-induced stresses during transport and handling, as well as its protective performance for the contents.</p>
<p>The loaded and sealed container is dropped from a defined height in free fall onto a rigid, non-yielding impact surface. For this purpose, INNOPROOF engineers developed a proprietary drop tower. The test enables targeted drops onto different areas of the container:</p>
<p> 	Flat drop: Impact flat on one of the side walls.<br />
 	Edge drop: Targeted impact on an edge.<br />
 	Corner drop: Direct impact on a corner (often the most critical stress).<br />
 <br />
Following the drop, the container is inspected for external damage (cracks, buckling, opening of closures), and the contents are examined for damage or leakage.</p>
<p>Der Beitrag <a href="https://innoproof.de/en/astm-d5276-drop-test-of-loaded-containers-by-free-fall/">ASTM D5276: Drop Test of Loaded Containers by Free Fall</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></description>
										<content:encoded><![CDATA[<p>The ASTM D5276 standard (“Standard Test Method for Drop Test of Loaded Containers by Free Fall”) describes a standardized test procedure for determining the impact resistance of filled shipping containers. Its objective is to evaluate the ability of packaging (such as boxes, crates, drums, or sacks) to withstand drop-induced stresses during transport and handling, as well as its protective performance for the contents.</p>
<p>The loaded and sealed container is dropped from a defined height in free fall onto a rigid, non-yielding impact surface. For this purpose, INNOPROOF engineers developed a proprietary drop tower. The test enables targeted drops onto different areas of the container:</p>
<ul>
<li>Flat drop: Impact flat on one of the side walls.</li>
<li>Edge drop: Targeted impact on an edge.</li>
<li>Corner drop: Direct impact on a corner (often the most critical stress).</li>
</ul>
<p>Following the drop, the container is inspected for external damage (cracks, buckling, opening of closures), and the contents are examined for damage or leakage.</p>
<p>Der Beitrag <a href="https://innoproof.de/en/astm-d5276-drop-test-of-loaded-containers-by-free-fall/">ASTM D5276: Drop Test of Loaded Containers by Free Fall</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></content:encoded>
					
		
		
			</item>
		<item>
		<title>ASTM F1829 – Glenoid shear test</title>
		<link>https://innoproof.de/en/astm-f1829-glenoid-shear-test-2/</link>
					<comments>https://innoproof.de/en/astm-f1829-glenoid-shear-test-2/#respond</comments>
		
		<dc:creator><![CDATA[Admin21]]></dc:creator>
		<pubDate>Tue, 29 Sep 2026 10:11:06 +0000</pubDate>
				<category><![CDATA[Testing Services]]></category>
		<category><![CDATA[Mechanical testing of shoulder replacements]]></category>
		<guid isPermaLink="false">https://innoproof.de/?p=1736</guid>

					<description><![CDATA[<p>ASTM F1829 describes a test method to determine the static shear force required to loosen modular glenoid components (insert and backing) of shoulder replacements. The test is suitable for metal joints, polymers, or composites and is used for design validation or comparison with other replacements. Wherever possible, the test should reflect the clinical use of the implant, meaning representative specimens of the finished product are tested and sterilized as required by the manufacturer.</p>
<p>For testing, specimens are fixed in the testing machine parallel to the implant axis. A vertical load is first applied from inferior to superior, and the loosening rate is recorded. The test is stopped when one of the following occurs:</p>
<p> 	The insert separates from the backing<br />
 	The applied load reaches its maximum and begins to decrease<br />
 	Gross deformation of the insert occurs without loosening</p>
<p>After this, a new insert is placed in the test fixture, and the test is repeated under the same conditions but applied from anterior to posterior. The test is stopped again as soon as one of the above cases occurs. The backing is visually inspected for damage after each test.</p>
<p>At least five equivalent samples should be tested, and the test can be conducted either in air at room temperature or under defined physiological conditions.</p>
<p>Der Beitrag <a href="https://innoproof.de/en/astm-f1829-glenoid-shear-test-2/">ASTM F1829 – Glenoid shear test</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></description>
										<content:encoded><![CDATA[<p><strong>ASTM F1829</strong> describes a test method to determine the static shear force required to loosen modular glenoid components (insert and backing) of shoulder replacements. The test is suitable for metal joints, polymers, or composites and is used for design validation or comparison with other replacements. Wherever possible, the test should reflect the clinical use of the implant, meaning representative specimens of the finished product are tested and sterilized as required by the manufacturer.</p>
<p>For testing, specimens are fixed in the testing machine parallel to the implant axis. A vertical load is first applied from inferior to superior, and the loosening rate is recorded. The test is stopped when one of the following occurs:</p>
<ul>
<li>The insert separates from the backing</li>
<li>The applied load reaches its maximum and begins to decrease</li>
<li>Gross deformation of the insert occurs without loosening</li>
</ul>
<p>After this, a new insert is placed in the test fixture, and the test is repeated under the same conditions but applied from anterior to posterior. The test is stopped again as soon as one of the above cases occurs. The backing is visually inspected for damage after each test.</p>
<p>At least five equivalent samples should be tested, and the test can be conducted either in air at room temperature or under defined physiological conditions.</p>
<p>Der Beitrag <a href="https://innoproof.de/en/astm-f1829-glenoid-shear-test-2/">ASTM F1829 – Glenoid shear test</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></content:encoded>
					
					<wfw:commentRss>https://innoproof.de/en/astm-f1829-glenoid-shear-test-2/feed/</wfw:commentRss>
			<slash:comments>0</slash:comments>
		
		
			</item>
		<item>
		<title>DIN EN 455-4 shelf life of gloves*</title>
		<link>https://innoproof.de/en/din-en-455-4-shelf-life-of-gloves/</link>
					<comments>https://innoproof.de/en/din-en-455-4-shelf-life-of-gloves/#respond</comments>
		
		<dc:creator><![CDATA[Admin21]]></dc:creator>
		<pubDate>Mon, 23 Mar 2026 15:50:21 +0000</pubDate>
				<category><![CDATA[Testing Services]]></category>
		<category><![CDATA[Mechanical testing of medical gloves]]></category>
		<guid isPermaLink="false">https://innoproof.de/?p=1284</guid>

					<description><![CDATA[<p>The gloves are stored in consumer packaging for the intended shelf life at a temperature specified by the manufacturer. Afterwards, the gloves are tested for watertightness according to EN 455-1 and for tensile strength according to EN 455-2. The integrity of the packaging and the suitability of the gloves for their intended use are also verified. The shelf life corresponds to the time at which all specified tests are still successfully passed.</p>
<p>Der Beitrag <a href="https://innoproof.de/en/din-en-455-4-shelf-life-of-gloves/">DIN EN 455-4 shelf life of gloves*</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></description>
										<content:encoded><![CDATA[<p>The gloves are stored in consumer packaging for the intended shelf life at a temperature specified by the manufacturer. Afterwards, the gloves are tested for watertightness according to <strong>EN 455-1</strong> and for tensile strength according to<strong> EN 455-2</strong>. The integrity of the packaging and the suitability of the gloves for their intended use are also verified. The shelf life corresponds to the time at which all specified tests are still successfully passed.</p>
<p>Der Beitrag <a href="https://innoproof.de/en/din-en-455-4-shelf-life-of-gloves/">DIN EN 455-4 shelf life of gloves*</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></content:encoded>
					
					<wfw:commentRss>https://innoproof.de/en/din-en-455-4-shelf-life-of-gloves/feed/</wfw:commentRss>
			<slash:comments>0</slash:comments>
		
		
			</item>
		<item>
		<title>DIN EN 455-2 physical properties of gloves*</title>
		<link>https://innoproof.de/en/din-en-455-2-physical-properties-of-gloves/</link>
					<comments>https://innoproof.de/en/din-en-455-2-physical-properties-of-gloves/#respond</comments>
		
		<dc:creator><![CDATA[Admin21]]></dc:creator>
		<pubDate>Mon, 23 Mar 2026 15:49:45 +0000</pubDate>
				<category><![CDATA[Testing Services]]></category>
		<category><![CDATA[Mechanical testing of medical gloves]]></category>
		<guid isPermaLink="false">https://innoproof.de/?p=1282</guid>

					<description><![CDATA[<p>According to DIN EN 455-2, the dimensions and the tensile strength of a medical disposable glove are determined. At least 13 samples per glove batch are used. For dimensional verification, the glove length and width are measured using a ruler and compared with the size table specified in the standard.</p>
<p>Depending on the intended use of the gloves, different dimensional requirements apply. For determination of tensile strength, dumbbell-shaped specimens are punched from the palm area of the gloves. The single wall thickness of the glove and the thickness of the dumbbell specimen are measured. The statically determined tensile force is multiplied by the ratio of single wall thickness to specimen thickness. The resulting tensile force value is then compared with the values specified in the standard.</p>
<p>Der Beitrag <a href="https://innoproof.de/en/din-en-455-2-physical-properties-of-gloves/">DIN EN 455-2 physical properties of gloves*</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></description>
										<content:encoded><![CDATA[<p><img decoding="async" class="alignright wp-image-327 size-medium" src="https://innoproof.de/wp-content/uploads/2025/12/EN-455-2-Zugversuch-182x300.jpg" alt="DIN EN 455-2 Physikalische Eigenschaften von Handschuhen*" width="182" height="300" srcset="https://innoproof.de/wp-content/uploads/2025/12/EN-455-2-Zugversuch-182x300.jpg 182w, https://innoproof.de/wp-content/uploads/2025/12/EN-455-2-Zugversuch-200x329.jpg 200w, https://innoproof.de/wp-content/uploads/2025/12/EN-455-2-Zugversuch.webp 400w" sizes="(max-width: 182px) 100vw, 182px" />According to <strong>DIN EN 455-2</strong>, the dimensions and the tensile strength of a medical disposable glove are determined. At least 13 samples per glove batch are used. For dimensional verification, the glove length and width are measured using a ruler and compared with the size table specified in the standard.</p>
<p>Depending on the intended use of the gloves, different dimensional requirements apply. For determination of tensile strength, dumbbell-shaped specimens are punched from the palm area of the gloves. The single wall thickness of the glove and the thickness of the dumbbell specimen are measured. The statically determined tensile force is multiplied by the ratio of single wall thickness to specimen thickness. The resulting tensile force value is then compared with the values specified in the standard.</p>
<p>Der Beitrag <a href="https://innoproof.de/en/din-en-455-2-physical-properties-of-gloves/">DIN EN 455-2 physical properties of gloves*</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></content:encoded>
					
					<wfw:commentRss>https://innoproof.de/en/din-en-455-2-physical-properties-of-gloves/feed/</wfw:commentRss>
			<slash:comments>0</slash:comments>
		
		
			</item>
		<item>
		<title>DIN EN 455-1 glove watertightness*</title>
		<link>https://innoproof.de/en/din-en-455-1-glove-watertightness/</link>
					<comments>https://innoproof.de/en/din-en-455-1-glove-watertightness/#respond</comments>
		
		<dc:creator><![CDATA[Admin21]]></dc:creator>
		<pubDate>Mon, 23 Mar 2026 15:49:00 +0000</pubDate>
				<category><![CDATA[Testing Services]]></category>
		<category><![CDATA[Mechanical testing of medical gloves]]></category>
		<guid isPermaLink="false">https://innoproof.de/?p=1280</guid>

					<description><![CDATA[<p>According to DIN EN 455-1, a water leak test can be performed to verify the watertightness of disposable gloves. Using a vertical tube, 1000 ml of water at a temperature between 15 °C and 35 °C is filled into the glove to be tested. The glove is immediately visually inspected for leaks. The glove is considered non-watertight as soon as water leakage occurs. After 2 to 3 minutes, the glove is visually inspected again.</p>
<p>Der Beitrag <a href="https://innoproof.de/en/din-en-455-1-glove-watertightness/">DIN EN 455-1 glove watertightness*</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></description>
										<content:encoded><![CDATA[<p><img decoding="async" class="alignright wp-image-326 size-full" src="https://innoproof.de/wp-content/uploads/2025/12/16_EN-455-1-2-4-Medical-glove-tests.jpg" alt="DIN EN 455-4 Haltbarkeit von Handschuhen* " width="274" height="389" />According to <strong>DIN EN 455-1</strong>, a water leak test can be performed to verify the watertightness of disposable gloves. Using a vertical tube, 1000 ml of water at a temperature between 15 °C and 35 °C is filled into the glove to be tested. The glove is immediately visually inspected for leaks. The glove is considered non-watertight as soon as water leakage occurs. After 2 to 3 minutes, the glove is visually inspected again.</p>
<p>Der Beitrag <a href="https://innoproof.de/en/din-en-455-1-glove-watertightness/">DIN EN 455-1 glove watertightness*</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></content:encoded>
					
					<wfw:commentRss>https://innoproof.de/en/din-en-455-1-glove-watertightness/feed/</wfw:commentRss>
			<slash:comments>0</slash:comments>
		
		
			</item>
		<item>
		<title>ASTM F2028 – Glenoid loosening test*</title>
		<link>https://innoproof.de/en/astm-f2028-glenoid-loosening-test/</link>
					<comments>https://innoproof.de/en/astm-f2028-glenoid-loosening-test/#respond</comments>
		
		<dc:creator><![CDATA[Admin21]]></dc:creator>
		<pubDate>Mon, 23 Mar 2026 15:45:07 +0000</pubDate>
				<category><![CDATA[Testing Services]]></category>
		<category><![CDATA[Mechanical testing of shoulder replacements]]></category>
		<guid isPermaLink="false">https://innoproof.de/?p=1266</guid>

					<description><![CDATA[<p>We offer a testing procedure to evaluate the fixation strength of the glenoid component in bone and its resistance to subluxation caused by cyclic movement (e.g., superior-inferior or anterior-posterior) of the humeral head against the edge of the glenoid, according to the ASTM F2028 standard. This test method can be applied to both cemented monolithic and modular glenoid components as well as uncemented reverse glenoid components.</p>
<p>The glenoid component is fixed in a bone substitute material using bone cement, and the humeral head component is placed into the glenoid component. An axial load is applied through the glenoid component while the humeral head is moved in opposite directions to determine the maximum displacement before dislocation. Then, the humeral head is cyclically moved at a speed of up to 50 mm/min for 100,000 cycles in both directions until 90% of the displacement at dislocation is reached. During the test, axial displacement (rocking) of the glenoid edges is recorded. The test can be conducted dry or in water at 37°C.</p>
<p>Der Beitrag <a href="https://innoproof.de/en/astm-f2028-glenoid-loosening-test/">ASTM F2028 – Glenoid loosening test*</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></description>
										<content:encoded><![CDATA[<p><img decoding="async" class="alignright wp-image-294 size-medium" src="https://innoproof.de/wp-content/uploads/2025/12/07_ASTM-F2028-Glenoid-loosening-shoulder-300x200.jpg" alt="ASTM F2028 Glenoid-Lockerungsprüfung* " width="300" height="200" srcset="https://innoproof.de/wp-content/uploads/2025/12/07_ASTM-F2028-Glenoid-loosening-shoulder-200x133.jpg 200w, https://innoproof.de/wp-content/uploads/2025/12/07_ASTM-F2028-Glenoid-loosening-shoulder-300x200.jpg 300w, https://innoproof.de/wp-content/uploads/2025/12/07_ASTM-F2028-Glenoid-loosening-shoulder-400x266.jpg 400w, https://innoproof.de/wp-content/uploads/2025/12/07_ASTM-F2028-Glenoid-loosening-shoulder.webp 500w" sizes="(max-width: 300px) 100vw, 300px" />We offer a testing procedure to evaluate the fixation strength of the glenoid component in bone and its resistance to subluxation caused by cyclic movement (e.g., superior-inferior or anterior-posterior) of the humeral head against the edge of the glenoid, according to the <strong>ASTM F2028</strong> standard. This test method can be applied to both cemented monolithic and modular glenoid components as well as uncemented reverse glenoid components.</p>
<p>The glenoid component is fixed in a bone substitute material using bone cement, and the humeral head component is placed into the glenoid component. An axial load is applied through the glenoid component while the humeral head is moved in opposite directions to determine the maximum displacement before dislocation. Then, the humeral head is cyclically moved at a speed of up to 50 mm/min for 100,000 cycles in both directions until 90% of the displacement at dislocation is reached. During the test, axial displacement (rocking) of the glenoid edges is recorded. The test can be conducted dry or in water at 37°C.</p>
<p>Der Beitrag <a href="https://innoproof.de/en/astm-f2028-glenoid-loosening-test/">ASTM F2028 – Glenoid loosening test*</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></content:encoded>
					
					<wfw:commentRss>https://innoproof.de/en/astm-f2028-glenoid-loosening-test/feed/</wfw:commentRss>
			<slash:comments>0</slash:comments>
		
		
			</item>
		<item>
		<title>ASTM F2203 linear dimensions*</title>
		<link>https://innoproof.de/en/astm-f2203-linear-dimensions/</link>
					<comments>https://innoproof.de/en/astm-f2203-linear-dimensions/#respond</comments>
		
		<dc:creator><![CDATA[Admin21]]></dc:creator>
		<pubDate>Mon, 23 Mar 2026 15:37:40 +0000</pubDate>
				<category><![CDATA[Testing Services]]></category>
		<category><![CDATA[Mechanical testing of packaging for medical devices]]></category>
		<guid isPermaLink="false">https://innoproof.de/?p=1246</guid>

					<description><![CDATA[<p>The dimensions of the packaging are determined using a ruler.</p>
<p>Der Beitrag <a href="https://innoproof.de/en/astm-f2203-linear-dimensions/">ASTM F2203 linear dimensions*</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></description>
										<content:encoded><![CDATA[<p>The dimensions of the packaging are determined using a ruler.</p>
<p>Der Beitrag <a href="https://innoproof.de/en/astm-f2203-linear-dimensions/">ASTM F2203 linear dimensions*</a> erschien zuerst auf <a href="https://innoproof.de/en">INNOPROOF GmbH</a>.</p>
]]></content:encoded>
					
					<wfw:commentRss>https://innoproof.de/en/astm-f2203-linear-dimensions/feed/</wfw:commentRss>
			<slash:comments>0</slash:comments>
		
		
			</item>
	</channel>
</rss>